Key Points Of Laboratory Magnetic Reactor Operation
Jun 25, 2023
Leave a message
Key points of laboratory magnetic reactor operation:
1. Reactors should be placed indoors. When multiple reactors are equipped, they should be placed separately. Each operating room should have an outlet directly leading to the outside or passage, and the equipment should be well ventilated.
2. When installing the lid of the magnetic reactor for the laboratory, the sealing surfaces between the reactor body and the reactor lid should be prevented from colliding with each other. Carefully place the reactor lid on the kettle body at a fixed position, and tighten the main When the nut is used, it must be tightened diagonally and symmetrically several times. The force should be uniform, and the lid of the reactor is not allowed to tilt to one side to achieve a good sealing effect.
3. At the connection between positive and negative nuts, only the positive and negative nuts are allowed to be rotated, and the two circular arc sealing surfaces must not be rotated relative to each other. When all nut pattern joints are assembled, lubricating oil should be applied. Gently turn the valve needle and press the cover tightly to achieve a good sealing effect. Turn the rotating body on the high-pressure magnetic reactor by hand to check whether the operation is flexible.
4. The controller should be placed flat on the operating table. The working environment temperature is 10-40°C, the relative humidity is less than 85%, and the surrounding medium does not contain conductive dust and corrosive gases. Check the movable parts on the panel and rear panel Check whether the connectors and fixed contacts are normal, pull off the upper cover, and check whether the contact of the connector is loose, whether there is damage or rust caused by improper transportation and storage. The laboratory reactor controller should be reliably grounded. Connect all wires, including the power cord , The electric furnace wire, motor wire, temperature sensor and speedometer wire between the controller and the reactor. Close the "power" air master switch on the panel, and the digital display should display. Set the various parameters on the digital display. Parameters (such as upper limit alarm temperature, working temperature, etc.) Then, press the "heating" switch, the electric furnace is turned on, and the indicator light on the "heating" switch is on at the same time. Adjust the "voltage adjustment" knob to adjust the heating power of the electric furnace. (for electric heating high-pressure magnetic reactor) press the "stirring" switch, the magnetic stirring motor is energized, and at the same time the indicator light on the "stirring" switch is on, slowly turn the "speed adjustment" knob to make the motor rotate slowly, and observe whether the motor is rotating forward or not. When there is no error, stop the machine and hang up the belt, and then restart.
5. After the operation is over, the laboratory reactor can be cooled naturally, cooled by water or placed on a stand for air cooling. After the temperature drops, the pressurized gas in the reactor can be released to reduce the pressure to normal pressure (the pressure gauge shows zero. ), then unscrew the main nut symmetrically and evenly, then remove the main nut, then carefully remove the kettle cover and place it on the bracket. After each operation, the residue on the reactor body and the magnetic reactor kettle cover should be removed. The main seal should be cleaned frequently and kept clean, and it is not allowed to wipe the magnetic reactor with hard objects or rough surfaces.