Pressure measurement process of stainless steel hydrogenation reactor

May 27, 2024

Leave a message

1. Use a conduit to connect the nitrogen bottle and the air inlet of the reactor, and tighten the relevant screws.
2. Open the main valve and partial pressure valve of the nitrogen bottle, adjust the pressure of the partial pressure valve to the pressure required for the experiment, then open the air inlet valve of the reaction kettle, and slowly fill it into the reaction kettle.
3. The pressure value displayed by the reactor is the same as the pressure set by the nitrogen bottle. When it no longer changes, close the reactor air inlet valve and the nitrogen bottle exhaust valve in sequence, record the pressure value displayed by the reactor, and observe the pressure after 30 minutes. Are there any changes?
The operating steps of the stainless steel hydrogenation reactor are as follows:
1) Before turning on the power switch of the console, set the stirring switch and speed regulating heating switch to zero to ensure normal startup.
2) Before feeding, rinse the inner wall of the kettle with distilled water, stir, rinse the mating surface, wipe cotton or silk cloth with ethanol, and slowly close the kettle lid to avoid hitting the mating surface.
3) When tightening the screws, you must sit in the correct position and use a twist wrench to apply cross-symmetrical and even force on the ground parts to avoid uneven force. Do not tighten the screws at once. Please apply force gradually and tighten symmetrically.
4) After tightening the screws, use a mechanical pump to evacuate, repeat nitrogen 2 to 4 times to discharge the air in the kettle, vacuum and introduce hydrogen, start stirring, and heat up. The operation is normal.
5) After the reaction is completed, stop stirring, extract the remaining hydrogen in the kettle, add nitrogen, vent, and loosen the screws to take samples.
6) The hydrogenation reactor is not resistant to strong acids, and strong acids such as hydrochloric acid, sulfuric acid, and nitric acid cannot be used in the reaction solution.
7) It is prohibited to produce sparks near the experimental site where small autoclaves are used, and it is prohibited to wear spiked shoes. When installing and using, please strictly follow the operating steps and instructions to avoid causing property and life losses to businesses and individuals.